A three-dimensional method of assessing the outcome of

orbital reconstruction after trauma

E. Brouns

∗

, M. Vehmeijer, N. Liberton, M. van Eijnatten,

T. Forouzanfar

Department of Oral and Maxillofacial Surgery/3D Innovationlab

VU University Medical Center, Amsterdam, The Netherlands

Background and Objective:

Orbital blow-out fractures are com-

monly a result of motor vehicle accidents, interpersonal violence,

or sports injuries and often require reconstructive surgery. A

plethora of different surgical approaches are currently used in

orbital floor reduction. These include transconjunctival, subcil-

iary, mid-lower eyelid, infraorbital and, more recently, endoscopic

transantral approaches. Some of the aforementioned treatment

strategies are combined with additional measures like intraopera-

tive imaging or navigation to enhance the accuracy hence surgical

outcome. The current treatment method at the VU University

Medical Center consists of a combination of virtual preopera-

tive planning and three-dimensional (3D) printing. This novel

approach has resulted in very good “subjective” clinical outcomes.

However, an “objective” method of assessing the outcome of

orbital reconstructions after trauma is still sought.

Methods:

To date, 15 patients have been operated using the afore-

mentioned method. In order to assess the overall accuracy and

the outcome of the orbital floor operation, virtually planned STL

models were compared with the postoperative STL models of

three patients. This presentation will describe all steps required

to objectively assess the outcome of orbital reconstruction after

trauma.

Findings:

The proposed 3D method demonstrated a strong agree-

ment between the virtually planned and the resulting orbital floor

reduction. This novel strategy offers a possible alternative to intra-

operative CT and navigation in the treatment of orbital floor

patients.

Conclusions:

Pre- and postoperative STL models can be used to

evaluate the accuracy of orbital floor reconstructions after trauma.

http://dx.doi.org/10.1016/j.ijom.2017.02.276

Magnetic resonance versus computed tomography: imaging

bony fractures of the orbit with view to surgical

reconstruction — a retrospective feasibility study

T. Cooper

∗

, B. Schmutz, E. Hsu, A. Lynham

University of Queensland, Queensland, Australia

Objective:

To examine the disparity between orbital fractures

imaged with magnetic resonance (MR) versus those imaged with

computed tomography (CT) using virtual segmentation and three-

dimensional (3D) modelling.

Methods:

Data from a previous study involving patients with

suspected unilateral orbital fracture recruited from Royal Bris-

bane and Women’s Hospital Maxillofacial Surgery clinics was

used. The primary predictor variable was the imaging technique

(MR versus CT) and the secondary predictor variable was type

of fracture (orbital blowout versus zygomaticomaxillary complex

[ZMC]). Outcome measurements were average deviation of frac-

ture margins and greatest deviation of fracture margins as defined

on MR and CT based virtual 3D models.

Findings:

Of the cases made available three were ZMC fractures

and seven were blowout orbital fractures (

n

= 10). Defect margins

mapped fromMR exhibited an average of 1.24 mm deviation from

CT overall and an average greatest deviation of 3.10 mm. Blowout

fractures (

n

= 7) exhibited an overall average deviation of 0.85 mm

and average greatest deviation of 2.01 mm. ZMC fractures (

n

= 3)

exhibited an overall average deviation of 2.14 mm and average

greatest deviation of 5.64 mm. There was a significant difference

between blowout and ZMC fractures average deviation (0.85 mm

versus 2.14 mm,

P

< 0.01).

Conclusion:

Orbital blowout fractures can be defined on MR with

clinically acceptable accuracy as evidenced by the overall average

deviation of 0.85 mm. The same cannot be said for ZMC frac-

tures. MR should be considered as an alternative imaging modality

for orbital blowout fracture when the risks of CT are clinically

significant.

http://dx.doi.org/10.1016/j.ijom.2017.02.277

Meta-analysis of orbitozygomatic complex trauma. The

cost–benefit relationship for the usefulness of eye care and

orthoptic screening

S. Crimi

∗

, A. Calvo, M. Sergiampietri, E. Nastro, F.S. De Ponte

Maxillofacial Unit, G.Martino Policlinic Hospital, University of

Messina, Italy

Objectives:

The aim of the study is to demonstrate the importance

of the ophthalmologic and orthoptic evaluation in the manage-

ment of the orbitozygomatic complex fractures in both stages:

immediately and in the postoperative follow-up.

Methods:

The ophthalmologic and orthoptic assessment at the

“Time 0” (acute) and distance can have indeed a double advantage:

from a clinical point of view to have objective data on the outcome

of the surgery and from an ethical point of view to improve the

“concordance” of the patient, showing him the outcome of the

situation pre- and postsurgery. Only in 2015, 81 patients have

been hospitalised, 74 of whom underwent surgery. In particular

40 fractures COMZ (54%), 23 fractures of the orbital floor (31%),

4 fractures interesting only the medial wall (5%), 1 NOE fracture

(2%), 6 fractures Le Fort (8%).

Results:

All patients were studied clinically and functionally with

Hess-Lancaster screen before and after the surgery. Seven patients

have been managed conservatively for the absence of symptoms

and signs of ocular movements involvement detected by Lancaster

screen. Three patients still present diplopia.

Conclusions:

A thorough clinical study of ophthalmologic and

orthoptics evaluation with Hess-Lancaster screen in the pre- and

postoperative appear so necessary both in the indication of surgical

timing, both in identifying the presence or onset of diplopia appear

useful in order to establish needs and timing of surgical treatment.

Advantage of Hess Lancaster screen is the capability to quantify

through numerical parameters the entity of ocular impairment and

help the clinician in order to treat and follow such patients in a

correct way.

http://dx.doi.org/10.1016/j.ijom.2017.02.278